Maquet Cardiopulmonary Ag: Medical Device Recall in 2018 - (Recall #: Z-1642-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

CARDIOHELP Gas Bottle Holder HKH 8880, Product Code/Part Number 70105.0901 Product Usage: The Cardiohelp Gas Bottle (or Cylinder) Holder HKH 8880 is an accessory for the CARDIOHELP System and required for inter-hospital patient transport in aircraft and road vehicles (mobile intensive care units and ambulances). Inter-hospital patient transport is transport of the patient between clinics, where the clinical environment is left (e.g., transport in vehicles such as ambulances or aircraft). The gas cylinder holder serves to hold an oxygen cylinder in aircraft.

Product Classification:

Class II

Date Initiated: April 4, 2018
Date Posted: May 16, 2018
Recall Number: Z-1642-2018
Event ID: 79811
Reason for Recall:

There is a possibility for the clamps of the gas bottle holder to crack or break when exposed to strong vibrations.

Status: Terminated
Product Quantity: 14 units (1 US, 13 International)
Code Information:

US: 98880-020 Foreign: 98880 - 002, 98880 - 004, 98880 - 005, 98880 - 006, 98880-009, 98880-010, 98880-011, 98880-012, 98880- 013, 98880-014, 98880-016, 98880-017, 98880-018

Distribution Pattern:

The product was distributed to the following US state: MO. The product was distributed to the following foreign countries: France, Germany, Switzerland, and Thailand.

Voluntary or Mandated:

Voluntary: Firm initiated