Maquet Cardiopulmonary Ag: Medical Device Recall in 2018 - (Recall #: Z-2161-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Aortic Arterial Cannula 24 Fr (O.D.); 23 cm (L); with 3/8-inch connector with LL, Part Number 701002267, Article Number A24-7107. For use in cardiopulmonary bypass

Product Classification:

Class II

Date Initiated: May 14, 2018
Date Posted: June 20, 2018
Recall Number: Z-2161-2018
Event ID: 80076
Reason for Recall:

This lot of arterial cannula may have been supplied with 3/8-width de-airing plugs that could break during the process of removal from the arterial cannula.

Status: Terminated
Product Quantity: 150 US
Code Information:

UDI CODE: 04037691239804, Lot Number: 92180017, 92183437 and 92184278

Distribution Pattern:

Worldwide distribution including US state of VA and Puerto Rico, Australia, Belgium, Brazil, Canada, Chile, Colombia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Iran, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Martinique, Mauritius, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated