Maquet Cardiovascular, LLC: Medical Device Recall in 2019 - (Recall #: Z-1499-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Maquet Getinge-BEQ-TOP 8013 ECC QUADROX iD Pack RF32 Material: 701047962

Product Classification:

Class II

Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1499-2019
Event ID: 82703
Reason for Recall:

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Status: Terminated
Product Quantity: N/A
Code Information:

Batch Numbers: 3000047630 3000050482 3000055980 3000060726 3000066813 3000068491 3000070813 3000074323

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated