Maquet Cardiovascular, LLC: Medical Device Recall in 2019 - (Recall #: Z-1526-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Maquet Getinge-BO-TOP 20705 SMALL PATIENT 1/4 Material: 701053486
Product Classification:
Class II
Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1526-2019
Event ID: 82703
Reason for Recall:
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Status: Terminated
Product Quantity: N/A
Code Information:
Batch Numbers: 3000048716 3000054549 3000061302 3000067224 3000070815 3000072509 3000076464
Distribution Pattern:
Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated