Maquet Cardiovascular, LLC: Medical Device Recall in 2019 - (Recall #: Z-1533-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Maquet Getinge-BEQ-TOP 32100 ADULT CUSTOM PAC Material:701054438
Product Classification:
Class II
Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1533-2019
Event ID: 82703
Reason for Recall:
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Status: Terminated
Product Quantity: N/A
Code Information:
Batch Numbers: 3000043291 3000044425 3000046811 3000047912 3000052203 3000053795 3000059174 3000062478 3000066197 3000067469 3000070703 3000074122 3000075896
Distribution Pattern:
Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated