Product Description:

Maquet Getinge-BEQ-TOP 38900 ECC PACK Material: 701056368

Product Classification:

Class II

Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1547-2019
Event ID: 82703
Reason for Recall:

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Status: Terminated
Product Quantity: N/A
Code Information:

Batch Numbers: 3000041396 3000044053 3000047317 3000050887 3000058131 3000060487 3000060723 3000062458 3000064724 3000071815 3000071817

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated