Maquet Cardiovascular, LLC: Medical Device Recall in 2019 - (Recall #: Z-1559-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Maquet Getinge-BEQ-TOP 24105 ECC PACK Material:701064757
Product Classification:
Class II
Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1559-2019
Event ID: 82703
Reason for Recall:
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Status: Terminated
Product Quantity: N/A
Code Information:
Batch Numbers: 3000051271 3000057632 3000062460 3000066202 3000069076 3000073425 3000076761 3000080703
Distribution Pattern:
Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated