Maquet Cardiovascular, LLC: Medical Device Recall in 2019 - (Recall #: Z-1579-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551

Product Classification:

Class II

Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1579-2019
Event ID: 82703
Reason for Recall:

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Status: Terminated
Product Quantity: N/A
Code Information:

Batch Numbers: 3000041542 3000046483 3000049065 3000055328 3000057682 3000060483 3000063712

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated