Maquet Cardiovascular, LLC: Medical Device Recall in 2019 - (Recall #: Z-1582-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Maquet Getinge-BEQ-T 8068 1/4" WITH BRIDGE Material: 701066883
Product Classification:
Class II
Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1582-2019
Event ID: 82703
Reason for Recall:
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Status: Terminated
Product Quantity: N/A
Code Information:
Batch Numbers: 3000041327 3000042148 3000042819 3000046732 3000054315 3000055642 3000062038 3000066689
Distribution Pattern:
Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated