Maquet Cardiovascular, LLC: Medical Device Recall in 2019 - (Recall #: Z-1658-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Maquet Getinge- (1) Material: 701063255R01 BEQ-TOP 25003 3/8 3/8 up to 7lpm (2) Material: 701063255R02 BEQ-TOP 25003 3/8 3/8 up to 7 lpm

Product Classification:

Class II

Date Initiated: April 18, 2019
Date Posted: June 5, 2019
Recall Number: Z-1658-2019
Event ID: 82703
Reason for Recall:

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Status: Terminated
Product Quantity: N/A
Code Information:

Batch Numbers: (1)3000070816, 3000072682 (2) 3000074736, 3000077083, 3000080988, 3000082175 ***Updated 3/26/20*** Extended: 3000087051

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated