Maquet Cardiovascular, LLC: Medical Device Recall in 2023 - (Recall #: Z-2272-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and chest surgery.

Product Classification:

Class II

Date Initiated: June 28, 2023
Date Posted: August 9, 2023
Recall Number: Z-2272-2023
Event ID: 92446
Reason for Recall:

The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.

Status: Ongoing
Product Quantity: 401 units
Code Information:

UDI-DI: 00607567700543

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.

Voluntary or Mandated:

Voluntary: Firm initiated