Maquet Cardiovascular Us Sales, Llc: Medical Device Recall in 2012 - (Recall #: Z-0160-2013)
See the recall detail below. You can also see other recalls from the same firm in 2012.
QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended for the use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The devices utilization period is limited to six hours.
Class II
It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.
All sizes and membrane types of QUADROX-i oxygenators, i.e., QUADROX-i(D) Adult, Small Adult, Pediatric and Neonatal with SOFTLINE and BIOLINE Coating; Catalogue numbers to include BE-HMOD XXXXX; BEQ-HMO XXXXX; BEQ-HMOD XXXX; BEQ-HMOD XXXXX; BO-VKMO XXXXXX; HMO XXXX; HMO XXXXX; HMOD XXXX; HMOD XXXXX; VKMO XXXX; VKMO XXXXX; X HMO XXXXX U; XHMOD 30000 U; X VKMO XXXXX U; and X 1895. QUADROX-iR with SOFTLINE and BIOLINE Coating - Catalogue numbers to include XVIVO HMO 70100 U. HLS module 5.0 and 7.0 in HLS Sets: Catalogue numbers to include BEQ-HLS 5050 and BEQ-HLS 7050. HLS module 5.0 and 7.0 in HIT HLS Sets: Catalogue numbers to include BO-HLS 5050 and BO-HLS 7050. Custom Tubing Packs containing oxygenators listed above: Catalogue numbers to include TOP XXXXX; TOP XXXXXns; TOP XXXX; BO-TOP XXXXX; BEQ TOP XXXXX; BSQ-TOP XXXX; and BSQ-TOP XXXXX. Note: an X was used to represent variations of catalogue numbers.
US Nationwide Distribution
Voluntary: Firm initiated