Maquet Cardiovascular Us Sales, Llc: Medical Device Recall in 2016 - (Recall #: Z-0348-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extra corporeal circulation and for blankets with which patients can be warmed or cooled. The HCU 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Product Classification:

Class II

Date Initiated: September 27, 2016
Date Posted: November 9, 2016
Recall Number: Z-0348-2017
Event ID: 75271
Reason for Recall:

There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.

Status: Terminated
Product Quantity: 305 units
Code Information:

HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653 HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.

Distribution Pattern:

Worldwide Distribution - US Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated