Maquet Medical Systems USA: Medical Device Recall in 2021 - (Recall #: Z-2140-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.
Class II
There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.
Distribution to US states of AL, CA, FL, IL, KS, MI, MO, NC, NE & OH.
Voluntary: Firm initiated