Maquet Medical Systems USA: Medical Device Recall in 2023 - (Recall #: Z-1640-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
QUADROX -i Neonatal, Oxygenator used during cardiopulmonary bypass Model Item no. HMO 10000-USA 701070412 HMO 11000-USA 701070416
Class II
1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.
Model Item no. UDI-DI HMO 10000-USA 701070412 4058863154473 HMO 11000-USA 701070416 4058863154558 All lots
Domestic distribution nationwide.
Voluntary: Firm initiated