Maquet Medical Systems USA: Medical Device Recall in 2023 - (Recall #: Z-1642-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Venous Hardshell Cardiotomy Reservoir, used during cardiopulmonary bypass Model Item No. VKMO 10000-USA 701070441 VKMO 11000-USA 701070445

Product Classification:

Class II

Date Initiated: May 18, 2023
Date Posted: June 7, 2023
Recall Number: Z-1642-2023
Event ID: 92003
Reason for Recall:

1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

Status: Ongoing
Product Quantity: 0 (US)
Code Information:

Model Item No. UDI-DI VKMO 10000-USA 701070441 4058863153841 VKMO 11000-USA 701070445 4058863153889 All lots

Distribution Pattern:

Domestic distribution nationwide.

Voluntary or Mandated:

Voluntary: Firm initiated