Maquet Medical Systems USA: Medical Device Recall in 2024 - (Recall #: Z-1831-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.

Product Classification:

Class II

Date Initiated: April 15, 2024
Date Posted: May 22, 2024
Recall Number: Z-1831-2024
Event ID: 94450
Reason for Recall:

The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.

Status: Ongoing
Product Quantity: 148 units (US: 5; OUS: 143)
Code Information:

Part No. 70104.8012 and 70104.8012R; UDI-DI: 04037691658384; Serial no. 90410021, 90410029, 90410165, 90410254, 90410255, 90410024, 90410025, 90410026, 90410028, 90410030, 90410031, 90410034, 90410035, 90410036, 90410043, 90410044, 90410045, 90410046, 90410047, 90410048, 90410049, 90410050, 90410051, 90410052, 90410053, 90410054, 90410055, 90410056, 90410057, 90410058, 90410059, 90410060, 90410061, 90410062, 90410063, 90410064, 90410065, 90410067, 90410068, 90410069, 90410070, 90410071, 90410072, 90410073, 90410074, 90410075, 90410076, 90410077, 90410078, 90410079, 90410080, 90410081, 90410082, 90410084, 90410085, 90410086, 90410087, 90410088, 90410089, 90410091, 90410093, 90410094, 90410095, 90410096, 90410097, 90410098, 90410099, 90410100, 90410101, 90410102, 90410103, 90410104, 90410105, 90410106, 90410107, 90410108, 90410109, 90410110, 90410111, 90410112, 90410113, 90410114, 90410115, 90410116, 90410117, 90410118, 90410119, 90410120, 90410121, 90410122, 90410123, 90410124, 90410125, 90410126, 90410127, 90410130, 90410131, 90410132, 90410133, 90410134, 90410135, 90410136, 90410137, 90410138, 90410139, 90410140, 90410141, 90410142, 90410143, 90410144, 90410146, 90410147, 90410148, 90410149, 90410151, 90410152, 90410153, 90410154, 90410155, 90410156, 90410157, 90410158, 90410159, 90410160, 90410161, 90410162, 90410163, 90410166, 90410167, 90410168, 90410169, 90410170, 90410172, 90410173, 90410174, 90410175, 90410176, 90410177, 90410178, 90410179, 90410180, 90410181, 90410182, 90410183, 90410184, 90410185, 90410186, 90410187.

Distribution Pattern:

US Nationwide distribution in the states of NJ & TX.

Voluntary or Mandated:

Voluntary: Firm initiated