Mc3 Inc: Medical Device Recall in 2024 - (Recall #: Z-1310-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 48135E
Class II
May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
GTIN:(1)10854916006871; (2)10854916006895 All Serial Numbers between 481022923 and 481024039
US Nationwide distribution.
Voluntary: Firm initiated