Mckesson Information Solutions LLC: Medical Device Recall in 2014 - (Recall #: Z-0093-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

Product Classification:

Class II

Date Initiated: June 25, 2014
Date Posted: October 29, 2014
Recall Number: Z-0093-2015
Event ID: 69380
Reason for Recall:

McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i

Status: Terminated
Product Quantity: 625 devices
Code Information:

McKesson Cardiology TM Hemo versions: 11.0HF5, 11.1, 11.1SP5, 11.1SP6, 11.1SP6HF1, 11.1SP6HF3, 12SP1, 12.1, 12.1HF1, 12.1HF2, 12.2, 12.2HF1, 12.2HF2, 12.2HF3, 12.2HF4.

Distribution Pattern:

Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated