Mechatronics USA: Medical Device Recall in 2023 - (Recall #: Z-1694-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
RR Mechatronics Starrsed Control Level N ESR Control 4.6 ml QRR049001- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026577
Class III
QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.
UDI-DI: (01)08719189137118(17)240222(10)QCAB6AN505 Lot Number/Exp Date: QCAB6AN505 24-02-2022
Worldwide distribution - US Nationwide distribution in the state of OH and the country of Netherlands.
Voluntary: Firm initiated