Med-Mizer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1541-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
Class II
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th
Model 762: 300000-300191 Model 763: 200000 Model 763A: 400000-40103
USA (nationwide)
Voluntary: Firm initiated