Med Tec Inc: Medical Device Recall in 2018 - (Recall #: Z-0671-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal X, Y, Z and yaw motions to provide 6-degree axis movement. The product includes an interface between the treatment planning system and the treatment table to direct the table motion based on coordinates supplied. These products connect or support a treatment table by use of various mounting plates.

Product Classification:

Class II

Date Initiated: November 8, 2017
Date Posted: February 28, 2018
Recall Number: Z-0671-2018
Event ID: 79182
Reason for Recall:

When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.

Status: Terminated
Product Quantity: 5 installations
Code Information:

MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.1 (M881920, M880610), MT6XSM1.7.2 (Lot M924170), MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3 (Lot M093990), MT6XSM1.7.3-1, MT6XSM1.7.3-3

Distribution Pattern:

US Distribution to Iowa and South Dakota. Shipped internationally to the Netherlands.

Voluntary or Mandated:

Voluntary: Firm initiated