Medefil Incorporated: Medical Device Recall in 2014 - (Recall #: Z-1149-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Heparin I.V. Flush Syringe is available is pre-filled polypropylene syringes, 100 units/mL, 3 mL fill in a 12 mL syringe.

Product Classification:

Class III

Date Initiated: February 19, 2014
Date Posted: March 12, 2014
Recall Number: Z-1149-2014
Event ID: 67562
Reason for Recall:

The bar code that is printed on the Master Carton label does not correspond to the product code fully displayed on the Master Carton label. The non-matching information can only be detected by using a bar code reader. The defective bar codes will be read as Product No. MIH-3335 ( Heparin I.V. Flush Syringe 100 units/mL; 5 mL fill in 12 mL Syringe) whereas this product is actually Product No. MIH-3333 (Heparin I.V. Flush Syringe 100 units/mL; 3 mL fill in 12 mL Syringe).

Status: Terminated
Product Quantity: 1,608 Master Cartons
Code Information:

Product Number: MIH-3333; NDC Number: 64253-333-33; Lot Numbers and Expiration Dates: Lot H114120N, Expires 12/15; Lot H113403N, Expires 11/15; Lot H113369N, Expires 10/15; Lot H113327N, Expires 8/15; Lot H113291N, Expires 7/15; Lot H113266N, Expires 6/15; Lot H113232N, Expires 5/15; Lot H113185N, Expires 3/15.

Distribution Pattern:

USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, SC, TX, UT, and WI, and the country of Canada.

Voluntary or Mandated:

Voluntary: Firm initiated