Medical Components, Inc dba MedComp: Medical Device Recall in 2018 - (Recall #: Z-1125-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Class II
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Lot Numbers: MCFF700, MCFB160, MKAG11O, MCDP590, MCDS140 & MCFSOSO
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
Voluntary: Firm initiated