Medical Components, Inc dba MedComp: Medical Device Recall in 2018 - (Recall #: Z-1271-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

8F Plastic Dignity¿ Low Profile CT Port W/Pre-Attached ChronoFlex¿ Polyurethane Catheter

Product Classification:

Class II

Date Initiated: November 22, 2017
Date Posted: April 4, 2018
Recall Number: Z-1271-2018
Event ID: 79445
Reason for Recall:

The kit was packaged with the incorrect guidewire. The label indicates the kit contains a marked guidewire. The kits were packaged with an unmarked guidewire.

Status: Terminated
Product Quantity: 24 Eaches
Code Information:

Lot/Serial/Model/Catalog Numbers/Exp. Date: Lot # MJKK420; serial # N/A, Catalog # MRDP80PLN, Exp. Date 2022-01-09; UDI#- 884908110449

Distribution Pattern:

CA, AR. AL

Voluntary or Mandated:

Voluntary: Firm initiated