Medical Positioning Inc: Medical Device Recall in 2014 - (Recall #: Z-2209-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations.

Product Classification:

Class II

Date Initiated: July 10, 2014
Date Posted: August 20, 2014
Recall Number: Z-2209-2014
Event ID: 68763
Reason for Recall:

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

Status: Terminated
Product Quantity: 2 beds
Code Information:

Serial number--803942

Distribution Pattern:

Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.

Voluntary or Mandated:

Voluntary: Firm initiated