Medical Positioning Inc: Medical Device Recall in 2014 - (Recall #: Z-2211-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Used to conduct syncope examinations.
Product Classification:
Class II
Date Initiated: July 10, 2014
Date Posted: August 20, 2014
Recall Number: Z-2211-2014
Event ID: 68763
Reason for Recall:
Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
Status: Terminated
Product Quantity: 1 bed
Code Information:
Serial number--803976
Distribution Pattern:
Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.
Voluntary or Mandated:
Voluntary: Firm initiated