Medicina Uk Ltd: Medical Device Recall in 2022 - (Recall #: Z-0111-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Medicina Oral Tip Syringe 2.5ml Code: OT25

Product Classification:

Class II

Date Initiated: July 4, 2022
Date Posted: October 26, 2022
Recall Number: Z-0111-2023
Event ID: 90849
Reason for Recall:

Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification

Status: Ongoing
Product Quantity: 20 boxes (2000 syringes)
Code Information:

UDI-DI:(01)05060278503431 Batch number: HM19026, HM19165, HM20019, HM20064, HM21061

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated