Medicina Uk Ltd: Medical Device Recall in 2022 - (Recall #: Z-0113-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Medicina Reusable Oral syringe 5ml Code: OTH05

Product Classification:

Class II

Date Initiated: July 4, 2022
Date Posted: October 26, 2022
Recall Number: Z-0113-2023
Event ID: 90849
Reason for Recall:

Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification

Status: Ongoing
Product Quantity: 26 boxes (2600 syringes)
Code Information:

UDI-DI: (01)05056115403491 (01)05056115403491(17)260700(10)HM21055 Batch number: HM18043, HM20020, HM21055

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated