Medicrea International: Medical Device Recall in 2020 - (Recall #: Z-0463-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
MEDICREA LIGAPASS TENSION PULLEY HANDLE XS- intended to provide thoracic and/or lumbar posterior stabilization for degenerative pathologies Ref : A08200200
Product Classification:
Class II
Date Initiated: October 8, 2020
Date Posted: November 25, 2020
Recall Number: Z-0463-2021
Event ID: 86657
Reason for Recall:
Tightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention ,
Status: Terminated
Product Quantity: 17 units US; 8 units OUS Expanded: 1 US and 3 OUS
Code Information:
Lot Numbers: US: 17E0550-1, 17E1089R/2, 18E0278, OUS: 19B0962, 19B0595, 19B0601 Expanded Recall: 11/3/20 17E0550/2, 19B0602, 19B0963 and 19B0964
Distribution Pattern:
CA Foreign:France, Belgium
Voluntary or Mandated:
Voluntary: Firm initiated