Medistim Asa: Medical Device Recall in 2025 - (Recall #: Z-0410-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No

Product Classification:

Class II

Date Initiated: September 26, 2025
Date Posted: November 5, 2025
Recall Number: Z-0410-2026
Event ID: 97707
Reason for Recall:

Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.

Status: Ongoing
Product Quantity: 14
Code Information:

Lot Code: Serial number; Model number 5042; MQC04001 5043; MQC04001 5044; MQC14001 5061; MQU14011 5062; MQU14011 5064; MQU14011 5065; MQU14011 5066; MQU14011 5067; MQC04001 5037; MQU14011 5073; MQV12011 5038; MQU14011 5045; MQU14011 5063; MQU14011

Distribution Pattern:

Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia.

Voluntary or Mandated:

Voluntary: Firm initiated