Medline Industries Inc: Medical Device Recall in 2016 - (Recall #: Z-2404-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.
Product Classification:
Class II
Date Initiated: May 11, 2016
Date Posted: August 17, 2016
Recall Number: Z-2404-2016
Event ID: 74664
Reason for Recall:
There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.
Status: Terminated
Product Quantity: 903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups)
Code Information:
Lot Numbers 122211502 122212515
Distribution Pattern:
Continental U.S.
Voluntary or Mandated:
Voluntary: Firm initiated