MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2022 - (Recall #: Z-1857-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T

Product Classification:

Class II

Date Initiated: August 3, 2022
Date Posted: October 5, 2022
Recall Number: Z-1857-2022
Event ID: 90722
Reason for Recall:

Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.

Status: Ongoing
Product Quantity: 136,080 units
Code Information:

a. UDI-DI (GTIN) case: (01)20888277703466; UDI-DI (GTIN) Each: (01)10888277703469, Lot Numbers 6922030011; b. UDI-DI (GTIN) case: (01)20888277703473; UDI-DI (GTIN) Each: (01)10888277703476, Lot Numbers 6921050011, 6921060011, 6921040012, 6921110011, 6921120011 6921120012, 6922010011, 6922010012, 6922010031, 6922020031, 6922030011.

Distribution Pattern:

Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Panama, Qatar, United Arab Emirates

Voluntary or Mandated:

Voluntary: Firm initiated