MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2023 - (Recall #: Z-1070-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Sterile Procedural Trays, labeled as the following: a. ORAL SURGERY PACK-LF b. EXTRACTION PACK c. ORAL SURGERY PACK d. DENTAL MINOR PACK e. ORAL

Product Classification:

Class II

Date Initiated: November 17, 2022
Date Posted: February 15, 2023
Recall Number: Z-1070-2023
Event ID: 91386
Reason for Recall:

Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

Status: Ongoing
Product Quantity: 174 cases (750 units)
Code Information:

a. ORAL SURGERY PACK-LF, Item Number: DYNJ20491N, Case UDI/GTIN: 40195327024582, Unit UDI/GTIN: 10195327024581, Lot Number: 22IDA002; b. EXTRACTION PACK, Item Number: DYNJ49329F, Case UDI/GTIN: 40195327059638, Unit UDI/GTIN: 10195327059637, Lot Number: 22HBR730; c. ORAL SURGERY PACK, Item Number: DYNJ52404G, Case UDI/GTIN: 40195327222612, Unit UDI/GTIN: 10195327222611, Lot Number: 22IDA539,22HDB737; d. DENTAL MINOR PACK, Item Number: DYNJ83140, Case UDI/GTIN: 40195327232956, Unit UDI/GTIN: 10195327232955, Lot Number: 22JBN386; e. ORAL, Item Number: DYNJ902377G, Case UDI/GTIN: 40195327224746, Unit UDI/GTIN: 10195327224745, Lot Number: 22KBG529,22JBO834;

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.

Voluntary or Mandated:

Voluntary: Firm initiated