MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2023 - (Recall #: Z-1080-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Sterile Procedural Trays, labeled as the following: a. UNIVERSAL PLASTIC b. KNEE ARTHROSCOPY PACK

Product Classification:

Class II

Date Initiated: November 17, 2022
Date Posted: February 15, 2023
Recall Number: Z-1080-2023
Event ID: 91386
Reason for Recall:

Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

Status: Ongoing
Product Quantity: 26 cases (104 units)
Code Information:

a. UNIVERSAL PLASTIC, Item Number: DYNJ04047C, Case UDI/GTIN: 40888277855060, Unit UDI/GTIN: 10888277855069, Lot Number: 22IBH860,22IBJ054; b. KNEE ARTHROSCOPY PACK, Item Number: DYNJ24784C, Case UDI/GTIN: 40889942732587, Unit UDI/GTIN: 10889942732586, Lot Number: 22JBM001;

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.

Voluntary or Mandated:

Voluntary: Firm initiated