Product Description:

Surgical convenience kits labeled as: CENTURION MEDICAL PRODUCTS IR INSTRUMENT TRAY, Reorder Number SUT10675

Product Classification:

Class II

Date Initiated: December 7, 2022
Date Posted: February 22, 2023
Recall Number: Z-1126-2023
Event ID: 91483
Reason for Recall:

The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.

Status: Ongoing
Product Quantity: 192 units
Code Information:

Reorder Number SUT10675, GTIN 40653160062845, Lot # 22HMG259

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated