MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2024 - (Recall #: Z-0053-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

Product Classification:

Class II

Date Initiated: August 28, 2024
Date Posted: October 16, 2024
Recall Number: Z-0053-2025
Event ID: 95338
Reason for Recall:

CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.

Status: Ongoing
Product Quantity: 72 kits
Code Information:

Lot numbers 24AMA209, exp. 2025-06-30; 24BMB025, exp. 2025-08-31; 23LMA745, exp. 2025-07-31; UDI-DI each 10195327384494, UDI-DI case-40195327384495

Distribution Pattern:

US Nationwide distribution in the state of CA.

Voluntary or Mandated:

Voluntary: Firm initiated