MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2024 - (Recall #: Z-0562-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
(1) Nursing Skills kit, SKU #EDUC05038B, Case UPC 40195327019878, Non-sterile, For Training Purposes Only, Not For Human Use.
Product Classification:
Class I
Date Initiated: October 16, 2023
Date Posted: January 3, 2024
Recall Number: Z-0562-2024
Event ID: 93327
Reason for Recall:
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
Status: Ongoing
Product Quantity: 172 kits
Code Information:
(1) EDUC05038B - Lot numbers 22IDA181, 22FDB543, and 22ADB550, case UPC40195327019878;
Distribution Pattern:
US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
Voluntary or Mandated:
Voluntary: Firm initiated