MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2024 - (Recall #: Z-1434-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Centurion retractor kits labeled as: a) STERILE 6-1/2" SENN RETRACTOR (ST7185), Product Code 66795; b) STERILE RAGNELL RETRACTOR (RR602), Product Code 67525; c) ST. SENN RETRACTOR 3 PRG BLUNT (ST7085), Product Code 37535

Product Classification:

Class II

Date Initiated: February 23, 2024
Date Posted: April 10, 2024
Recall Number: Z-1434-2024
Event ID: 94136
Reason for Recall:

Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

Status: Ongoing
Product Quantity: 760 units
Code Information:

a) 66795, UDI/DI 50653160061180 (case) 00653160061185 (each), Lot Numbers: 2023030190, 2023052390, 2023092590; b) 67525, UDI/DI 10653160158639 (case) 00653160158632 (ease), Lot Numbers: 2023032490; c) 37535, UDI/DI 10653160169048 (case) 00653160169041 (each), Lot Numbers: 2023061490

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Panama, Canada.

Voluntary or Mandated:

Voluntary: Firm initiated