MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2024 - (Recall #: Z-3111-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO BRONCH PACK, Pack Number DYKE1355; b) ENDO BASIC KIT, Pack Number DYKE1356; c) ENDOSCOPY KIT, Pack Number DYKE1771C

Product Classification:

Class II

Date Initiated: June 27, 2024
Date Posted: September 18, 2024
Recall Number: Z-3111-2024
Event ID: 95128
Reason for Recall:

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Status: Ongoing
Product Quantity: 6257 units
Code Information:

a) DYKE1355: UDI/DI case 40889942274193, UDI/DI each 10889942274192; Lot Numbers: 23LBB043, 24BBJ327, 24CBS751; b) DYKE1356: UDI/DI case 40889942274186, UDI/DI each 10889942274185; Lot Numbers: 23LBD062, 24ABA646, 24BBJ328, 24CBG449, 24DBE106, 24EBG238; c) DYKE1771C: UDI/DI case 40195327331994, UDI/DI each 10195327331993; Lot Numbers: 24ABX292

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Voluntary or Mandated:

Voluntary: Firm initiated