MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2024 - (Recall #: Z-3113-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR BREAST, Pack Number DYNJ908104B; b) PLASTIC MAJOR, Pack Number DYNJ908106B; c) ABDOMINOPLASTY W/LIPO SUCTION, Pack Number DYNJ909946; d) PLASTIC, Pack Number DYNJ909951
Class II
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
a) DYNJ908104B: UDI/DI case 40195327507115, UDI/DI each 10195327507114, Lot Numbers: 23LBT869, 24BBQ730; b) DYNJ908106B: UDI/DI case 40195327507139, UDI/DI each 10195327507138, Lot Numbers: 24ABB087, 24CBM739, 24EBH234; c) DYNJ909946: UDI/DI case 40195327512652, UDI/DI each 10195327512651, Lot Numbers: 23LBD934, 24DBB806; d) DYNJ909951: UDI/DI case 40195327512904, UDI/DI each 10195327512903, Lot Numbers: 23LBD932, 24ABP740
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Voluntary: Firm initiated