MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2025 - (Recall #: Z-0725-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
KIT,SPINE LH Convenience kit, SKU DYNJ906448A.
Product Classification:
Class II
Date Initiated: November 6, 2024
Date Posted: January 1, 2025
Recall Number: Z-0725-2025
Event ID: 95806
Reason for Recall:
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
Status: Ongoing
Product Quantity: 160 kits
Code Information:
UDI-DI: 10193489408508(each), 40193489408509(case); Lot Number: 21EBG271 21CBS664 20LBO883 20WBE707
Distribution Pattern:
US Nationwide. Canada.
Voluntary or Mandated: