MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2025 - (Recall #: Z-1430-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

namic convenience kits labeled as: MTS,COLOR,PML,KIT,-,FRENCH HOSPITAL PG; medical convenience kit, REF 600605710

Product Classification:

Class II

Date Initiated: February 12, 2025
Date Posted: April 2, 2025
Recall Number: Z-1430-2025
Event ID: 96367
Reason for Recall:

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

Status: Ongoing
Product Quantity: 100 units
Code Information:

UDI/DI 10193489070279 (each),20193489070276 (case), Lot number 0000141266

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia

Voluntary or Mandated:

Voluntary: Firm initiated