MEDLINE INDUSTRIES, LP - Northfield: Medical Device Recall in 2025 - (Recall #: Z-2199-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301
Class II
SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of disinfection of the luer access valve, potentially resulting in patient infection/sepsis.
UDI-DI 00363807103016 (ea) 10363807103013 (box) 20363807103010 (case) Lots 3144609 3144857 3144886 3144967 3145090 3145186
Worldwide distribution - US Nationwide and the countries of Canada and Panama.
Voluntary: Firm initiated