Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1013-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.
Class I
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
1) REF MDR107003E, UDI/DI 40080196321002, All lots; 2) REF MDR107003E-4, UDI/DI 40080196321002, All lots
US, CA, PA, GUAM, VI
Voluntary: Firm initiated