Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1014-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO

Product Classification:

Class I

Date Initiated: November 26, 2025
Date Posted: January 28, 2026
Recall Number: Z-1014-2026
Event ID: 98185
Reason for Recall:

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Status: Ongoing
Product Quantity: 22565 units
Code Information:

UDI/DI 10080196546992, All lots

Distribution Pattern:

US, CA, PA, GUAM, VI

Voluntary or Mandated:

Voluntary: Firm initiated