Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1090-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Product Classification:

Class II

Date Initiated: December 16, 2025
Date Posted: January 21, 2026
Recall Number: Z-1090-2026
Event ID: 98197
Reason for Recall:

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Status: Ongoing
Product Quantity: 12 units
Code Information:

UDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated