Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1197-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

Medline medical procedure convenience kits labeled as: 1) RR-GYN LAPAROSCOPY PACK-LF, Kit SKU DYNJ0395022O; 2) PELVISCOPY PACK-LF, Kit SKU DYNJ0415776Q; 3) DIVA PACK, Kit SKU DYNJ04811O; 4) DA VINCI PROSTATE/COLPOPEXY, Kit SKU DYNJ44864S; 5) MH ROBOTICS PACK, Kit SKU DYNJ50880L; 6) DELNOR LAPAROSCOPIC GYN PACK, Kit SKU DYNJ82244B; 7) ROBOT PACK, Kit SKU DYNJ82761B; 8) GENERAL ROBOTICS PACK, Kit SKU DYNJT4964.

Product Classification:

Class II

Date Initiated: December 12, 2025
Date Posted: February 4, 2026
Recall Number: Z-1197-2026
Event ID: 98226
Reason for Recall:

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

Status: Ongoing
Product Quantity: 1,928 total
Code Information:

Medline Kit SKU DYNJ0395022O: UDI/DI 10195327634827 (EA) 40195327634828 (CS), Lot Number 25KME682; Medline Kit SKU DYNJ0415776Q: UDI/DI 10198459566806 (EA) 40198459566807 (CS), Lot Number 25KMH125; Medline Kit SKU DYNJ04811O: UDI/DI 10198459440953 (EA) 40198459440954 (CS), Lot Number 25KMD772; Medline Kit SKU DYNJ44864S: UDI/DI 10195327127626 (EA) 40195327127627 (CS), Lot Number 25LMA115; Medline Kit SKU DYNJ50880L: UDI/DI 10198459388002 (EA) 40198459388003 (CS), Lot Number 25LMA516; Medline Kit SKU DYNJ82244B: UDI/DI 10198459344053 (EA) 40198459344054 (CS), Lot Number 25KMF328; Medline Kit SKU DYNJ82761B: UDI/DI 10198459039232 (EA) 40198459039233 (CS), Lot Number 25LMA391; Medline Kit SKU DYNJT4964: UDI/DI 10198459460098 (EA) 40198459460099 (CS), Lot Number 25KMJ341.

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated