Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1313-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
Intubation Tray (Adult) containing Suction Catheter Kits SKU DYNJAA245B
Class II
A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
Kit lots 25IBC172 25HBJ413 UDI (ea) 10198459480317 UDI (case) 40198459480318
Distribution US nationwide and Canada.
Voluntary: Firm initiated