Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1384-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number: DYNJ905567A

Product Classification:

Class II

Date Initiated: January 7, 2026
Date Posted: February 25, 2026
Recall Number: Z-1384-2026
Event ID: 98329
Reason for Recall:

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Status: Ongoing
Product Quantity: 551 units
Code Information:

1) DYNJ0185290G, UDI-DI: 10193489378054(each), 40193489378055(case), Lot Number: 22LBQ336; 2) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 24BBN429; 3) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23KBI089; 4) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23HBG596; 5) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBL644; 6) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBG152; 7) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23EBH550; 8) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23CBQ546; 9) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22NBF780; 10) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22LBC300; 11) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBP453; 12) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBM263; 13) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBF686; 14) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22CBY841; 15) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBR457; 16) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBA253; 17) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21IBH675; 18) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21GBE802; 19) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21EBE777

Distribution Pattern:

Worldwide distribution - US Nationwide.

Voluntary or Mandated:

Voluntary: Firm initiated